Reports 1-1 of 1 Clear search Modify search
DAQ (DAQ)
gouaty, masserot - 10:18 Tuesday 08 February 2011 (28776) Print this report
sampling frequency of Pr_B1_d2(3)_ACp_40K increased to 80K
Today around 08:50 UTC we changed the sampling frequency of the channels Pr_B1_d2_ACp_40K and Pr_B1_d3_ACp_40K to put it at 80kHz. We also changed the cutting frequency of the Butterworth. It is now at 78000 Hz in Moni_Det1.cfg
These changes have been made in order to try to observe the OMC 28kHz signal with these photodiodes, when the OMC is locked.
Once this measurement is performed we will go back to the original situation for these two channels
Comments to this report:
masserot - 14:37 Tuesday 08 February 2011 (28784) Print this report
To take in account this size increase, I ve changed the FDOUT_SHM size from 5MBytes to 6 MBytes. The FbMoniCb server was restarted to take in account this modification
Search Help
×

Warning

Error

The present report has been modified outside this window. Please check for its integrity in the main page.

Refreshing this page will move this report into drafts.

×