========================================================== Candidate mechanical resonance modes ========================================================== Instrumented baffle : WI Triaxial accelerometer position : BOTTOM Triaxial accelerometer axis : Z Reference mono-axial accelerometer position : R3 (horizontal, on tower vessel) Quiet acquisition : SIG0061 Injection acquisition : SIG0062 Candidate resonance selection criteria -------------------------------------- Peak prominence > 4.0 Coherence > 0.50 Phase rotation > 60 deg Freq [Hz] TF Mag. Prom. Coh. Phase Rot. [deg] Width Comment ----------------------------------------------------------------------------------------------- 78.25 14.771 9.54 0.638 94.0 1.35 - 100.25 10.647 9.38 0.892 112.7 3.05 - 163.00 5.109 4.96 0.672 116.6 2.06 - 179.25 5.466 4.20 0.799 122.1 2.27 - 244.75 11.840 11.75 0.785 145.2 2.73 - 272.50 9.774 8.36 0.838 142.8 2.89 - 355.00 18.336 18.20 0.718 167.2 1.74 - 359.00 6.225 4.13 0.957 118.7 2.56 - 426.25 5.930 4.70 0.896 74.5 4.90 - 440.00 7.138 6.27 0.705 161.2 4.71 - 473.75 10.226 10.15 0.970 118.9 6.70 -