========================================================== Candidate mechanical resonance modes ========================================================== Instrumented baffle : WI Triaxial accelerometer position : BOTTOM Triaxial accelerometer axis : Y Reference mono-axial accelerometer position : R1 (horizontal, below baffle) Quiet acquisition : SIG0057 Injection acquisition : SIG0058 Candidate resonance selection criteria -------------------------------------- Peak prominence > 4.0 Coherence > 0.50 Phase rotation > 60 deg Freq [Hz] TF Mag. Prom. Coh. Phase Rot. [deg] Width Comment ----------------------------------------------------------------------------------------------- 100.25 11.934 11.63 0.800 129.9 3.05 - 146.50 30.180 26.73 0.656 221.2 6.74 Check phase rotation 169.75 31.140 28.95 0.772 149.9 2.55 - 355.50 4.283 4.10 0.927 161.7 2.25 - 440.25 5.320 4.00 0.837 69.1 4.06 - 491.25 5.372 4.13 0.930 178.1 4.70 - 505.75 10.528 10.21 0.764 103.7 4.70 - 587.00 18.681 16.86 0.631 215.1 2.38 - 597.75 14.729 10.37 0.863 145.6 5.97 - 664.25 14.204 12.52 0.854 144.4 4.16 - 678.00 8.883 6.51 0.895 165.6 5.66 - 684.50 13.610 5.15 0.667 105.3 1.52 - 689.50 26.262 15.36 0.889 187.8 1.74 - 794.00 8.202 7.77 0.813 242.4 5.40 Check phase rotation 810.50 6.468 5.90 0.737 156.1 3.28 - 820.25 5.924 4.56 0.799 104.3 3.71 - 882.25 6.317 5.57 0.623 67.7 3.69 - 944.50 6.563 5.12 0.920 81.3 8.71 - 976.75 7.446 7.27 0.780 121.5 7.10 -