========================================================== Candidate mechanical resonance modes ========================================================== Instrumented baffle : WI Triaxial accelerometer position : BOTTOM Triaxial accelerometer axis : X Reference mono-axial accelerometer position : R1 (horizontal, below baffle) Quiet acquisition : SIG0057 Injection acquisition : SIG0058 Candidate resonance selection criteria -------------------------------------- Peak prominence > 4.0 Coherence > 0.50 Phase rotation > 60 deg Freq [Hz] TF Mag. Prom. Coh. Phase Rot. [deg] Width Comment ----------------------------------------------------------------------------------------------- 123.00 6.681 4.93 0.684 102.2 1.43 - 140.75 37.783 31.12 0.879 149.2 3.46 - 146.50 93.367 90.18 0.650 153.6 2.76 - 272.75 5.374 4.59 0.673 197.3 3.09 - 423.75 15.581 12.64 0.560 229.8 6.24 Check phase rotation 440.00 6.975 5.52 0.835 70.5 4.03 - 473.75 6.419 6.03 0.964 133.9 7.37 - 522.50 8.318 7.08 0.816 502.4 6.13 Check phase rotation 540.00 7.985 6.10 0.552 111.1 2.77 - 562.00 11.996 8.88 0.634 164.2 2.64 - 578.00 25.114 22.71 0.537 140.3 4.99 - 587.00 28.510 26.37 0.637 205.3 2.30 - 594.25 26.084 17.23 0.791 74.5 1.96 - 620.50 12.496 10.98 0.924 89.7 7.82 - 637.75 13.295 9.64 0.543 172.1 2.47 - 647.50 19.630 16.97 0.527 106.0 2.51 - 674.25 7.473 6.69 0.835 90.4 4.96 - 683.50 11.845 7.48 0.886 86.8 2.04 - 689.25 13.627 9.40 0.913 177.8 3.01 -