========================================================== Candidate mechanical resonance modes ========================================================== Instrumented baffle : WI Triaxial accelerometer position : TOP Triaxial accelerometer axis : Y Reference mono-axial accelerometer position : R3 (horizontal, on tower vessel) Quiet acquisition : SIG0053 Injection acquisition : SIG0054 Candidate resonance selection criteria -------------------------------------- Peak prominence > 4.0 Coherence > 0.50 Phase rotation > 60 deg Freq [Hz] TF Mag. Prom. Coh. Phase Rot. [deg] Width Comment ----------------------------------------------------------------------------------------------- 68.25 14.187 14.13 0.679 118.1 1.50 - 78.00 6.362 6.03 0.578 93.8 1.48 - 100.00 5.600 5.53 0.861 90.5 2.00 - 244.75 6.414 5.61 0.860 82.4 1.92 - 272.50 18.169 17.04 0.802 148.5 2.66 - 355.00 26.919 26.90 0.619 162.1 2.29 - 421.00 13.715 13.52 0.646 128.4 3.88 - 440.25 7.204 6.42 0.654 162.2 3.94 - 505.50 4.312 4.07 0.899 118.9 6.88 -