========================================================== Candidate mechanical resonance modes ========================================================== Instrumented baffle : WI Triaxial accelerometer position : TOP Triaxial accelerometer axis : Y Reference mono-axial accelerometer position : R2 (vertical) Quiet acquisition : SIG0051 Injection acquisition : SIG0052 Candidate resonance selection criteria -------------------------------------- Peak prominence > 4.0 Coherence > 0.50 Phase rotation > 60 deg Freq [Hz] TF Mag. Prom. Coh. Phase Rot. [deg] Width Comment ----------------------------------------------------------------------------------------------- 68.50 15.225 12.38 0.651 160.2 1.70 - 78.00 7.464 7.12 0.603 99.7 1.61 - 100.00 5.489 5.39 0.896 99.9 1.88 - 129.75 11.779 11.58 0.754 112.8 2.95 - 141.75 8.833 6.76 0.824 88.7 1.75 - 169.25 9.245 8.16 0.640 115.0 3.25 - 178.25 6.975 5.86 0.631 227.1 4.07 Check phase rotation 244.50 8.946 8.24 0.672 91.6 1.72 - 272.25 13.232 13.22 0.844 132.9 2.87 - 355.50 5.908 5.75 0.766 152.5 1.98 - 508.75 6.221 5.75 0.848 121.4 2.94 -