========================================================== Candidate mechanical resonance modes ========================================================== Instrumented baffle : WI Triaxial accelerometer position : TOP Triaxial accelerometer axis : Y Reference mono-axial accelerometer position : R1 (horizontal, below baffle) Quiet acquisition : SIG0055 Injection acquisition : SIG0056 Candidate resonance selection criteria -------------------------------------- Peak prominence > 4.0 Coherence > 0.50 Phase rotation > 60 deg Freq [Hz] TF Mag. Prom. Coh. Phase Rot. [deg] Width Comment ----------------------------------------------------------------------------------------------- 68.25 12.717 12.71 0.652 112.2 1.61 - 100.00 6.064 6.00 0.840 103.8 1.86 - 118.50 5.392 5.20 0.560 77.7 1.50 - 141.50 9.760 9.65 0.790 124.3 2.19 - 147.50 8.463 4.28 0.531 75.0 1.41 - 158.25 9.307 7.80 0.632 65.2 1.41 - 170.25 5.096 4.03 0.880 219.8 3.82 - 273.00 13.358 11.06 0.742 142.8 2.28 - 421.25 7.436 6.15 0.599 165.9 6.19 - 440.25 4.438 4.05 0.865 128.4 5.41 -