========================================================== Candidate mechanical resonance modes ========================================================== Instrumented baffle : NI Triaxial accelerometer position : BOTTOM Triaxial accelerometer axis : X Reference mono-axial accelerometer position : R3 (horizontal, on tower vessel) Quiet acquisition : SIG0049 Injection acquisition : SIG0050 Candidate resonance selection criteria -------------------------------------- Peak prominence > 4.0 Coherence > 0.50 Phase rotation > 60 deg Freq [Hz] TF Mag. Prom. Coh. Phase Rot. [deg] Width Comment ----------------------------------------------------------------------------------------------- 57.75 6.388 4.11 0.614 70.7 1.10 - 64.25 9.055 5.77 0.601 109.0 2.32 - 80.50 9.275 7.09 0.515 118.5 1.53 - 122.75 19.043 13.57 0.508 86.6 1.24 - 129.25 24.474 22.72 0.759 120.6 3.18 - 137.75 21.173 12.80 0.642 84.1 1.50 - 179.75 5.234 4.74 0.667 78.7 6.80 - 198.75 7.037 6.40 0.821 117.0 3.96 - 211.25 4.837 4.02 0.543 110.9 4.03 - 253.50 4.893 4.26 0.635 143.0 4.70 - 351.50 30.227 29.77 0.958 147.1 4.03 - 361.75 14.663 10.96 0.893 101.6 3.29 - 376.50 8.690 4.96 0.962 121.9 4.96 - 399.00 42.304 42.24 0.765 100.4 3.37 - 407.25 23.857 20.59 0.843 447.4 12.64 Check phase rotation 444.25 18.732 17.03 0.855 139.4 3.91 - 458.00 9.844 7.02 0.825 127.5 2.74 - 468.00 17.075 16.54 0.750 235.3 5.97 Check phase rotation 508.25 33.054 32.39 0.970 109.6 4.91 - 524.50 24.854 20.28 0.930 80.9 3.19 - 686.00 6.868 6.49 0.678 136.9 4.73 - 901.75 5.121 4.86 0.903 121.5 10.17 -