========================================================== Candidate mechanical resonance modes ========================================================== Instrumented baffle : NI Triaxial accelerometer position : BOTTOM Triaxial accelerometer axis : X Reference mono-axial accelerometer position : R1 (horizontal, below baffle) Quiet acquisition : SIG0043 Injection acquisition : SIG0044 Candidate resonance selection criteria -------------------------------------- Peak prominence > 4.0 Coherence > 0.50 Phase rotation > 60 deg Freq [Hz] TF Mag. Prom. Coh. Phase Rot. [deg] Width Comment ----------------------------------------------------------------------------------------------- 80.50 7.202 4.72 0.563 140.9 1.89 - 129.25 26.972 23.19 0.809 118.2 3.14 - 143.00 11.620 8.48 0.502 125.6 2.48 - 211.25 6.345 5.89 0.562 136.8 2.80 - 315.00 4.419 4.28 0.566 119.3 2.91 - 352.25 34.077 34.47 0.950 101.9 4.07 - 361.75 11.299 8.41 0.909 119.7 2.43 - 415.25 19.819 17.91 0.968 175.0 12.32 - 425.75 12.356 8.00 0.887 65.6 2.77 - 444.50 13.548 12.21 0.895 171.9 4.87 - 471.25 31.431 30.54 0.753 188.0 4.05 - 479.25 21.044 11.95 0.539 162.3 3.43 - 509.00 70.552 70.49 0.918 119.8 3.51 - 545.50 7.621 5.94 0.908 66.9 5.25 - 602.50 20.088 17.11 0.761 477.1 11.07 Check phase rotation 607.50 15.447 4.90 0.927 126.6 2.57 -