========================================================== Candidate mechanical resonance modes ========================================================== Instrumented baffle : NI Triaxial accelerometer position : TOP Triaxial accelerometer axis : X Reference mono-axial accelerometer position : R3 (horizontal, on tower vessel) Quiet acquisition : SIG0038 Injection acquisition : SIG0040 Candidate resonance selection criteria -------------------------------------- Peak prominence > 4.0 Coherence > 0.50 Phase rotation > 60 deg Freq [Hz] TF Mag. Prom. Coh. Phase Rot. [deg] Width Comment ----------------------------------------------------------------------------------------------- 80.25 10.755 10.48 0.555 112.0 1.84 - 105.50 7.406 4.62 0.891 82.0 1.79 - 134.00 21.969 18.46 0.717 241.1 7.07 Check phase rotation 228.50 6.252 5.03 0.721 140.2 5.13 - 295.00 9.743 7.72 0.810 172.1 2.28 - 328.00 30.998 26.21 0.895 122.7 3.05 - 340.00 58.957 51.85 0.950 192.1 4.10 - 345.50 98.924 90.70 0.643 99.6 1.31 - 365.75 93.799 78.12 0.907 218.0 4.52 - 371.50 68.954 44.05 0.836 108.9 2.08 - 386.25 106.530 106.48 0.542 120.2 3.12 - 411.25 25.626 8.32 0.858 104.4 6.81 - 443.25 52.397 50.36 0.892 233.9 21.92 Check phase rotation 452.00 11.821 7.90 0.839 126.8 3.14 - 458.25 11.181 8.90 0.918 78.6 2.19 - 468.75 39.558 34.11 0.977 65.4 5.51 - 480.00 48.884 31.49 0.778 78.9 1.96 - 495.75 79.555 78.48 0.890 160.6 6.20 - 506.00 36.844 28.20 0.688 173.2 3.65 - 518.75 15.186 8.36 0.572 322.6 2.87 Check phase rotation 543.25 33.331 29.30 0.754 76.1 4.76 - 552.75 10.640 8.27 0.788 62.1 4.25 - 571.25 6.593 4.25 0.635 107.1 3.77 - 581.75 13.386 9.64 0.859 65.4 3.29 - 608.00 68.713 67.23 0.950 93.4 12.17 - 626.50 50.807 48.87 0.577 201.3 3.14 - 651.25 16.294 15.12 0.899 125.8 2.90 - 728.00 5.774 5.56 0.860 179.6 5.36 - 903.75 8.965 8.49 0.877 60.6 9.76 - 946.25 4.931 4.50 0.556 138.1 17.19 -