========================================================== Candidate mechanical resonance modes ========================================================== Instrumented baffle : NI Triaxial accelerometer position : TOP Triaxial accelerometer axis : Z Reference mono-axial accelerometer position : R2 (vertical) Quiet acquisition : SIG0036 Injection acquisition : SIG0037 Candidate resonance selection criteria -------------------------------------- Peak prominence > 4.0 Coherence > 0.50 Phase rotation > 60 deg Freq [Hz] TF Mag. Prom. Coh. Phase Rot. [deg] Width Comment ----------------------------------------------------------------------------------------------- 57.50 13.938 13.17 0.568 66.0 1.36 - 80.50 8.857 6.97 0.533 170.5 2.05 - 131.25 6.110 5.52 0.855 170.2 3.78 - 222.50 6.578 6.25 0.714 145.5 2.59 -