========================================================== Candidate mechanical resonance modes ========================================================== Instrumented baffle : NI Triaxial accelerometer position : TOP Triaxial accelerometer axis : Y Reference mono-axial accelerometer position : R2 (vertical) Quiet acquisition : SIG0036 Injection acquisition : SIG0037 Candidate resonance selection criteria -------------------------------------- Peak prominence > 4.0 Coherence > 0.50 Phase rotation > 60 deg Freq [Hz] TF Mag. Prom. Coh. Phase Rot. [deg] Width Comment ----------------------------------------------------------------------------------------------- 80.50 10.882 6.91 0.578 127.5 1.75 - 131.25 56.321 56.29 0.909 85.1 4.50 - 222.25 8.914 7.47 0.663 129.5 3.27 - 314.75 6.162 5.58 0.621 119.2 2.28 - 328.00 8.484 6.97 0.922 91.4 3.94 - 339.50 28.897 27.92 0.989 122.7 5.43 - 345.00 9.947 6.23 0.820 117.3 1.65 - 351.00 10.962 8.63 0.971 161.8 6.46 - 370.00 8.237 5.75 0.854 87.0 2.44 - 402.25 10.767 6.83 0.731 75.5 3.22 - 407.25 18.581 13.86 0.801 106.5 1.71 - 428.75 23.343 21.50 0.728 142.3 15.67 - 446.25 16.525 7.94 0.776 74.6 2.56 - 458.00 14.053 11.13 0.688 124.9 2.31 - 468.25 11.317 7.80 0.893 181.8 2.35 - 495.25 6.417 5.24 0.936 134.3 7.18 - 505.50 5.628 4.36 0.619 354.9 6.03 Check phase rotation 549.25 28.398 25.46 0.628 87.2 6.89 - 564.50 16.434 10.52 0.890 87.0 3.69 - 613.75 22.803 20.26 0.839 70.9 6.34 - 624.00 20.892 14.69 0.642 179.7 5.25 - 646.50 10.535 7.94 0.970 119.5 7.78 -