========================================================== Candidate mechanical resonance modes ========================================================== Instrumented baffle : NI Triaxial accelerometer position : TOP Triaxial accelerometer axis : Y Reference mono-axial accelerometer position : R1 (horizontal, below baffle) Quiet acquisition : SIG0041 Injection acquisition : SIG0042 Candidate resonance selection criteria -------------------------------------- Peak prominence > 4.0 Coherence > 0.50 Phase rotation > 60 deg Freq [Hz] TF Mag. Prom. Coh. Phase Rot. [deg] Width Comment ----------------------------------------------------------------------------------------------- 131.00 65.430 65.35 0.892 121.8 4.00 - 182.50 6.388 6.20 0.582 183.0 6.74 - 295.00 8.951 8.48 0.756 176.8 2.72 - 314.75 6.804 6.35 0.612 120.6 1.89 - 328.25 7.237 5.98 0.937 90.7 3.07 - 339.50 18.605 16.56 0.974 152.1 5.35 - 366.00 10.306 7.27 0.941 82.3 4.89 - 446.25 8.283 4.24 0.932 91.1 8.72 - 469.00 16.869 16.32 0.890 313.1 4.38 Check phase rotation 481.75 11.314 8.97 0.653 88.7 5.71 - 495.75 8.837 6.36 0.943 82.5 5.47 - 523.00 8.057 6.83 0.970 63.8 4.49 - 565.25 10.541 7.80 0.959 79.7 4.12 - 602.00 47.706 16.27 0.850 83.9 2.03 - 611.00 48.413 48.26 0.916 198.8 17.41 - 624.50 23.781 17.56 0.898 164.6 5.37 - 640.00 20.650 13.85 0.690 156.4 4.24 - 666.00 9.524 4.52 0.921 68.3 2.52 - 669.75 11.894 7.75 0.904 66.1 3.69 - 767.00 9.980 9.56 0.929 89.1 3.56 - 855.00 6.113 5.99 0.853 72.9 5.29 -