*** 28 November 2025, Tapping in Central Building**** 2 minutes at each location 9:28:44 SDB1 south flange (121 Hz and many) 9:33:47 cryotrap lato W DET viewport (also 315 Hz and many) 9:38:14 cryotrap vessel (only above 400Hz, peak 501Hz, 843 Hz) 9:41:57 SR flange East (much less, 501Hz, 840 Hz) (120Hz is always a bit excited fluctuating, also when not tapping. just moving around) 9:46:00 SR flange West (120 Hz) 9:50:00 BS East flange (nothing) 9:52:35 BS South flange (nothing) (glitch at 9:52:50) 9:57:00 PR North flange, at the POP flange (only 730 Hz, and something around 100Hz = 93 Hz 105 Hz) 10:01:40 PR South flange (nothing) 10:05:00 Cryotrap INJ Vessel (nothing) 10:08:24 SIB1 North flange (530-545 broad or double structure) 10:13:05 PR North flange (nothing) 10:16:48 link PR-BS (nothing) 10:28:37 SIB1 West flange (NOTHING) 10:36:52 NI north flange (165 Hz rises and disappears with a delay) 10:39:30 After almost 2 minutes it rises again, with a delay of about 1 minute 10:42:05 NI south flange -- unlock new lock 16:03:00 WI West Hrflange (115 Hz fluctuates but might be unrealted) (glitch at 16:04:46) the 115 seems went away about 1m after inside DLAB (note that there are still 2 acc on SDB2 chamber, one on south and one on east side) 16:22:46 B1 viewport (su flangia lunga)(26 AND 64) glitch at 16:23:50 (stop tapping) 16:26:40 B1 in proximity of viewport (directly on flange on link pipe) (26.5 fluctuating and 66 Hz) 16:30:34 B1p viewport (large 115-120 Hz,200, 322 and many) (26.5 Hz and 20-30 Hz keeps fluctuating since we are around SDB1\2) 16:33:40 glitch 16:34:57 Hartman on flange in proximity of viewport (only a little 66 Hz) 16:41:00 Hartmann viewport (66 breathes a little) 16:43:40 B5 flange (similar to effect seen on B1 flange more, effect above 400 Hz seems more intense, also 101 Hz,190 Hz) 16:49:06 EQB1 door (951 Hz narrow) 17:10:55 SPRB "North" flange (nothing evident) inside SIB2 clean room 17:29:30 (glitch) 17:30:38 SIB2 beam dump flange (nothing.. 120 Hz narrow is breathing but seems unrelated) 17:34:27 SIB1 beam dump (EAST) (peaks 386 and 480 Hz and 500-700Hz, 820 Hz)