May 11 2023 - Maria, Irene quiet 21:14:43 UTC to 31:31:00 UTC NOTA: hrec looks very noisy while LSC_DARM seems ok **** test sweep with different speed NEB sweep 5V 30-100 Hz 300s 21:31:00 UTC - 21:37:34 UTC (a few missing data at the beginning) can see something in DARM sweep 5V 30-100 Hz 1200s 21:39:18 UTC - 22:00:00 UTC glitch a circa 21:40 UTC can see it in DARM sweep 5V 30-100 Hz 600s 22:02:16 UTC - 22:15:32 UTC glitch at about 22:05:40 can see it in DARM **** test linearity by injecting lines with different levels NEB using automatic script of Catalina lines 36, 56, 115 300s 300s 300s amplitude 5 9 9 22:37:35 to 22:58 (first 300s are quiet) MagneticLine_NOISE_NEB-1367879871.txt amplitude 2.5 4.5 4.5 23:02:36 to 23:23 (first 300s are quiet) MagneticLine_NOISE_NEB-1367881374.txt ok lines are visible in hrec CEB sweep 10-500 9V 14400s started at 23:33:00