This afternoon, we performed a new HWS measurement after optimizing the position of the actuator on the base platform during the activities of yesterday afternoon and this morning. The HWS-DET acquisition started at 13.06 UTC (20230615T1505) in NI single bounce and B5 drift control disabled.
At around 13.13 UTC, the PA actuator has been switched ON (WF13) at around 11 W (4.331 A, 8 V).
The PA actuator ellipsoidal beam projected on the NI TM appeared clearly in the map (see an example in fig 1).
At 14.03 UTC the PA actuator has been switched off and, at 14.29 UTC, the HWS acquisition has been stopped.
After that, we tried to perform a new measurement with the ‘smile’ alignment mask (fig. 2) in front of the heater.
A new HWS-DET acquisition started at 14.57 UTC (20230615T1657) and the PA actuator has been switched ON at 13.03 UTC. The mask produces some ‘structures’ in the HWS maps not clearly recognizable with respect to the expected one (fig.3).
Thus, tomorrow morning, we will continue to roughly align the actuator without the mask in front of the heater by using the picomotors.