Reports of 58472
On-call intervention (General)
Oncall-system - 13:57 Tuesday 23 April 2024 (64073) Print this report
On-call intervention

The following report has been submitted to the On-call interface.

On-call events -> Air Conditioning

Title: Water leakage warning.

Author(s): andreazzoli

Called at: 00:55, 19-04-2024, by: Alarm or monitoring system
Remote intervention: Started: ; Ended:
On-site intervention: Started: 08:43, 19-04-2024; Ended: 09:20, 19-04-2024
Status: Resolved
Operator when issue resolved: Sposito

Details:

I received four calls from the control room between 00:51 and 01:15.
I did not hear the calls.
At 08:02 I called the control room and spoke to the operator.
Soldani Davide had intervened during the night.
I decided to intervene for a short check to verify what had happened and if possible take action in order to correct the problem.

* Note that any files attached to this report are available in the On-call interface.

AdV-DAQ (Calibration)
verkindt - 11:41 Tuesday 23 April 2024 (64069) Print this report
Hrec connected on FbmFE

Today, around 8h55 UTC, I changed the config of Hrec to connect it to FbmFE instead of FbmAlp and use the science mode flag DQ_META_ITF_Mode_FE
instead of DQ_META_ITF_Mode and use the lock index flag META_ITF_LOCK_index_FE instead of META_ITF_LOCK_index.
Since 20/04 8h30 UTC, this new configuration was tested on HrecN test process and showed a decrease of the mean latency and an independence
with respect to the latency glitches present on FbmAlp (see plot1).
To do this connection of Hrec to FbmFE, I had to install a new version of Hrec code v5r13 in virgoApp.

Images attached to this report
AdV-DAQ (Calibration)
verkindt - 11:24 Tuesday 23 April 2024 (64071) Print this report
BNS range correlations with DCP of BS, optical gain and hrec/hinj at 77.5 Hz

I already made this logbook entry as 64034, but it disappeared. So I put it here again:

Over the long lock that started in the night of 18/04 to 19/04, we can see a correlated trend between the BNS range which starts dropping after 19/04 10h UTC
and the behaviour of the DCP frequency estimated by Hrec for the BS mirror and also the behaviour of the modulus of hrec/hinj at 77.5 Hz.
On the same plot, we can see that after 19/04 10h UTC, the optical gain estimated by Hrec is fluctuating more (as well as the BNS range)
and that the variations of the optical gain (estimated with the calibration lines around 70 Hz) are well correlated with the variations of hrec/hinj modulus at 77.5 Hz
over le ss than an hour time scale.
The trends over those 18 hours of lock may be linked to the etalon scan.
The larger fluctuations of BNS range, optical gain or hrec/hinj after 19/04 10h UTC have an other explanation.
Finding it would allow to improve the stability of the BNS range and of the bias hrec/hinj.

Images attached to this report
AdV-DET (Optical, mechanical and electronic components for benches (optics and camera mounts, fast shutter, motors,..))
mwas, carbognani - 11:13 Tuesday 23 April 2024 (64070) Print this report
OMC slow shutter driver replacement

Today between 8:00 UTC and 9:00 UTC we have done two tests to investigate the issue of the OMC slow shutter with a broken stop.

Stop and start the driver. This did not resolve the issue, the shutter when opening still runs without stopping when arriving to the end. We have stopped it manually after 1 minute.

Replace the driver with a spare, serial number FT4ZNKQD. The configuration of the rotation process was updated with the new serial number. The issue remained the same. The translation stage of the shutter still doesn't stop when closing.

All these actions were performed with DET_MAIN in pause and the interferometer in single bounce.

In conclusion the driver is not the issue, and probably the in-vacuum translation stage has one of the two stops that is broken.

IT infrastructure (Computing)
kraja, cortese - 9:18 Tuesday 23 April 2024 (64068) Print this report
Firmware upgrade on underlying virtual infrastructure
During this maintenance window we will perform some necessary upgrades on the underlaying virtual infrastructure where the fileserver are hosted.
This operation is totally transparent for the hosted systems on the infrastructure, and should not impact the fileservers at all.
Virgo Runs (O4b)
menzione - 6:54 Tuesday 23 April 2024 (64066) Print this report
Operator Report - Night shift

Upon arrival I found ITF just relocked and in Science Mode.
Unlocks:
- 23:19 UTC (?). Relocked at first attempt. 
- 01:32 UTC probably due to the seismic activity in Taiwan. Relockeed at second attempt at 03:36 UTC. 

Guard Tour (UTC)
21:06 - 21:34
23:08 - 23:34
01:04 - 01:30
03:15 - 03:45
 

https://gracedb.ligo.org/superevents/S240422ed/view/

Images attached to this report
AdV-COM (AdV commissioning (1st part) )
casanueva, derossi, boldrini, bersanetti - 23:10 Monday 22 April 2024 (64065) Print this report
ALS noise

Today the lock acquisition was very troublesome due an excess of noise on the West Green beam. We tried to check the coupling of the fibers on the DAQ room, but it didn't solve the problem. Then we went to the end building and tried to improve the fiber coupling by moving the fibers in the rack. We found a position where the noise was reduced. Still there is some green power missing, so tomorrow some more systematic alignment work will be done, in order to better align the green in the SHG box. After this action the lock acquisition worked properly.

Virgo Runs (O4b)
Sposito - 22:56 Monday 22 April 2024 (64061) Print this report
Operator Report - Afternoon shift

Today, I found ITF unlocked with WI and WE guardian opened. I closed the guardian on both and realigned the mirrors to have back the flashes on B8, P. Ruggi used the dumper on WI Y to reduce the excitations of the mirror after the opening of the controls. I then aligned the CITF and restarted the locking acquisition. 15:30-15:40 UTC The WE guardian opened again. I properly closed and restored the position of the WE mirror that went off. M. Boldrini, J. Casanueva, D. Bersanetti, and M. Was are working on the ITF recovery. We discovered some problems, the first one was an issue with the WE ALS green (fixed by J. Casanueva and M. Boldrini with C. De Rossi from remote), second problem was with Ch e Das WI by mistake a flipper was left closed, I re-opened at 19:05UTC with the guidance of I. Nardecchia and the third problem was on PR_TX gain adjustments (D. Bersanetti and M. Boldrini). We had an 18:26 UTC earthquake in 6.3 Taiwan and an 18:32 UTC earthquake in 6.4 Taiwan. 20:50 UTC ITF back in SCIENCE. 

Images attached to this report
AdV-ISC (Commissioning up to first full interferometer lock)
boldrini - 21:31 Monday 22 April 2024 (64064) Print this report
Comment to PR_TX gain adjustmen (64060)

Yesterday's situation has been restored: PR_TX is reset to 2.2 in CARM_NULL_1F and is set to 2.2 as well in LOW_NOISE_1.

The gain change in LOW_NOISE_1 is kept in Metatron's code, but is currently doing nothing.

AdV-TCS (CO2 laser projector)
nardecchia, sposito - 21:28 Monday 22 April 2024 (64063) Print this report
WI thermal camera flip mirror closed

Around 19:00 LT, I discovered that the WI flip mirror used to acquire the thermal camera images was closed. This means that the CH and DAS were blocked on the bench since last night.

Gianmatteo re-opened it at 19.05 UTC.

AdV-ISC (Commissioning up to first full interferometer lock)
boldrini, bersanetti - 17:28 Monday 22 April 2024 (64060) Print this report
PR_TX gain adjustmen

A few of today's unlocks, after the wind settled down, have been caused by low gain oscillation for PR_TX (Fig.1).

Increasing the gain in CARM_NULL_1F caused an unlock soon after the transition, so we introduced a gain change in ACQUIRE_LOW_NOISE_1 (line 5232 and 5275 in ITF_LOCK.py).

The modifications can easily be reverted or made transparent (by simply setting this gain to the same value in CARM_NULL_1F and LOW_NOISE_1) if need be.

Images attached to this report
Comments to this report:
boldrini - 21:31 Monday 22 April 2024 (64064) Print this report

Yesterday's situation has been restored: PR_TX is reset to 2.2 in CARM_NULL_1F and is set to 2.2 as well in LOW_NOISE_1.

The gain change in LOW_NOISE_1 is kept in Metatron's code, but is currently doing nothing.

Virgo Runs (O4b)
berni - 15:02 Monday 22 April 2024 (64054) Print this report
Operator Report - Morning shift

6:00 UTC: ITF found in BAD_WEATHER due to high wind and sea activity.

At 10:13 UTC I removed the mode BAD_WEATHER because the related flags WIND and SEA on the DMS were not red anymore but unfurtunately the lock acquisition continued to fail at various state; Maddalena checked from the control room.

After a check we discovered a change in the WI CH power that happend last night thus, at 10:30 UTC, I set TROUBLESHOOTING and we contacted Ilaria to restore the good value of the powers. This problem was not reported on the DMS thus a readjustment of the thresholds is needed.

After this change the interferometer continued to unlock also due to a further worsening of the weather conditions: strong storm with lightning and thunder.

The commissioning team present in control room is investigating at the problems.

 

ALS

As reported in the previous shift the DMS is reporting a problem with WE ALS Laser; this morning I informed the experts by mail. At 12:00 UTC I did not receive any answer thus we decided to contact the ALS on-call; unfurtunately he is not available to look at the problem.

 

SUSP

At around 12:50 UTC WE and WI LC were opened by the guardian after an unlock. I tried to close the loops but the new GUI interface did not work; for this reason I contacted the SUSP on-call. Investigation on the problem in progress from control room.

Images attached to this report
AdV-DAQ (Data collection)
cortese, kraja, dibiase - 14:47 Monday 22 April 2024 (64059) Print this report
Comment to DAQ - TolmFrameBuilder unable to send theirs frames (64039)

These events are likely correlated  to the burst of jobs that have been automatically submitted to the HTCondor farm by the DQR and detchar pipelines some time before, triggered by several H1 GstLAL alerts.

Between 4:20 UTC and 6:20 UTC about 1600 of these jobs have run causing a peak of more than 150MB/s nfs read traffic and consequent overload  of the fs01 fileserver where /virgoLog is hosted.

Since RTPCs still mount /virgo, /virgoApp and /cvmfs via NFS from fs01 instead of CVMFS, they may have been impacted by the corresponding slowdown which however has not provoked any visible errors at the Operating System level.

 

Detector Characterisation (Broadband noise)
Paoletti, Boschi - 13:52 Monday 22 April 2024 (64058) Print this report
Comment to yesterday fly-over (64027)
Images attached to this comment
Detector Characterisation (Broadband noise)
ruggi - 12:30 Monday 22 April 2024 (64057) Print this report
Good BNS range in recent data

Sometimes we can find data with BNS range stable above 56 Mpc for several minutes. This is the best level we can reach since several weeks. In march we had something better: for instance, on March 8th the range was 58.9 Mpc, averaged over 10 min.

In order to understand where those 2 Mpc have been lost, the two sensitivities are compared in fig 1. In order to understand better, the two cumulative BNS ranges have been computed and subtracted (fig 2).

It is clear enough that the mystery noise is not involved: a change of its level should be well visible in the region 80 - 120 Hz, where instead the sensitivity is equal. The bigger drop is below 80 Hz, both because an increase of the broadband noise and the presence of new structures. If we want to reach the psychological threshold of 60 Mpc, this region needs to be better studied.

Images attached to this report
On-call intervention (General)
Oncall-system - 10:30 Monday 22 April 2024 (64056) Print this report
On-call intervention

The following report has been submitted to the On-call interface.

On-call events -> Air Conditioning

Title: water leak in COB officies conditioning pipe

Author(s): soldani

Called at: 01:15, 20-04-2024, by: Magazzu
Remote intervention: Started: ; Ended:
On-site intervention: Started: 01:34, 20-04-2024; Ended: 01:54, 20-04-2024
Status: Resolved
Operator when issue resolved: Magazzu

Details:

At the request of the In shift-Operator A.Magazzu', I reached the site to stop the water leak from the air conditioning pipe for the offices on the ground floor of the COB.
The relevant shutters were closed and the company in charge of maintenance was contacted, which this morning (22/04) is already working to replace the affected section of pipe (whose installation dates back to 1996) because its state of conservation does not allow a repair that gives a guarantee of reliability over time.
If there are no unforeseen events, the restoration work will be completed the same day.

* Note that any files attached to this report are available in the On-call interface.

On-call intervention (General)
Oncall-system - 9:28 Monday 22 April 2024 (64055) Print this report
On-call intervention

The following report has been submitted to the On-call interface.

On-call events -> Injection System

Title: pmc pzt corrections high

Author(s): derossi

Called at: 06:00, 22-04-2024, by: Menzione
Remote intervention: Started: 06:00, 22-04-2024; Ended: 06:10, 22-04-2024
On-site intervention: Started: ; Ended:
Status: Resolved
Operator when issue resolved: Menzione

Details:

I was called by the operator because the flag PMC PZT was red.
We had set the flag red when the PMC PZT corrections become larger than 4,5V because in this conditions the lock would probably not last more than a few hours and it could be worth to manually unlock the PMC to reset the corrections (during the lock the corrections are likely to increase until 7V, causing the PMC unlock). Since is not something which prevents the ITF to lock we changed the color of the flag in the DMS: yellow from 4,5V to 6,5 and red starting from 6.5V.

* Note that any files attached to this report are available in the On-call interface.

Virgo Runs (O4b)
menzione - 6:53 Monday 22 April 2024 (64053) Print this report
Operator Report - Night shift

Upon arrival I found ITF relocking after the earthquake.
Relocking in progress but ITF very noisy due to the high wind and sea activity increasing (plot3). The bad weather conditions prevent a stable lock for the whole shift.

ALS
I noticed that after the intervention at WE on ALS laser amplifier, the Green Power and the IR_PSTAB are not in a good working point (plot2).

SBE
After one of the unlock events SNEB Horizontal loop opened. Properly closed.

TCS
During a relock the CH TCS_CO2..REL3 got stuck at 1. I act ON Up via TTL_CO2.

Guard Tour (UTC)
21:06 - 21:34
22:55 - 23:25
01:04 - 01:50
03:07 - 03:39 

Oncall events

ISYS
(22-04-2024 04:00 - ) From remote
Status: On-going
Description: PSL_PMC_PZT out of range (plot1). Normally after an unlock it resets but not this night (?)
DMS comment: PMC_PZT_likely_to_drift_away.

Images attached to this report
Virgo Runs (O4b)
Sposito - 23:01 Sunday 21 April 2024 (64051) Print this report
Operator Report - Afternoon shift

Today, I found ITF in SCIENCE. 18.30 - 18:42 UTC I started the scheduled DF_DAILY calibration. 18:43 UTC ITF back in SCIENCE. 20:11 UTC An earthquake (5.5 Mexico) caused ITF to unlock. I realigned the CITF and restarted the locking acquisition. 20:59 UTC ITF unlocked at LOW_NOISE_2.

Sub-system reports

Other
VPM
18:43 UTC restarted CALI after a crash

Images attached to this report
Detector Characterisation (Broadband noise)
Paoletti - 22:22 Sunday 21 April 2024 (64052) Print this report
Comment to yesterday fly-over (64027)

" So probably the helicopter respected the NOTAM. "

I guess not

Images attached to this comment
Detector Characterisation (Broadband noise)
Boschi - 19:22 Sunday 21 April 2024 (64046) Print this report
Comment to yesterday fly-over (64027)

Nice idea. However the original restrictions - still enforced, as reported in the attached text of the forbidden areas of the latest release of the AIP publication (https://www.enav.it/servizi-online) - are cylinders of 2000 ft = 610 m of height but radius of 0.3 nautical miles = 1823 ft = 556 m. So probably the helicopter respected the NOTAM. 

Images attached to this comment
Interesting data (Interesting data)
Paoletti, Romero. - 16:05 Sunday 21 April 2024 (64049) Print this report
PC_1 magnetic pulsation
Images attached to this report
Non-image files attached to this report
Environmental Monitoring (Environmental Monitoring)
Paoletti, Fiori, Tringali. - 15:18 Sunday 21 April 2024 (64048) Print this report
Comment to Test of automated injections using ENV_MAIN (63714)
Images attached to this comment
Virgo Runs (O4b)
berni - 14:58 Sunday 21 April 2024 (64047) Print this report
Operator Report - Morning shift

ITF in Science mode for all the shift.

 

 

Guard tours (time in UTC)

From 6:00 to 6:30; from 8:00 to 8:30; from 10:00 to 10:30; from 12:00 to 12:30.

Images attached to this report
Virgo Runs (O4b)
amagazzu - 7:31 Sunday 21 April 2024 (64045) Print this report
Operator Report - Night shift

Upon reaching the site, I found the ITF locked in LOW_NOISE_3_SQZ and in Science Mode. At 22:12 UTC the Squeezing disengaged, bringing the ITF back in LOW_NOISE_3. ITF back in LOW_NOISE_3_SQZ from 22:14 UTC.
ITF left locked.

Guard Tour (UTC):
21:30 - 22:00
23:30 - 00:00
1:38 - 2:07
3:07 - 3:18

Sub-system reports

DAQ
The process ELE_QGBT was not working correctly, it was reported as not working on DMS and in Orange state on VPM. At 4:13 UTC I restarted the process to solve the issue.

Images attached to this report