Reports of 125 Clear search Modify search
AdV-PSL (Installation on site)
chaibi, soulard, turconi, derossi, chiummo - 14:30 Monday 14 March 2022 (55158) Print this report
Master laser moved back in the atrium

Yesterday we moved back the master laser (ML) in the atrium. In the first photo you can see the installation of the ML in the atrium's rack. The beam fibered path is: ML box --> pig-tailed EOM --> 300mW fiber AMP (S/N 609) --> laser bench

Here below are the details of the activities related to this installation.

We first blocked the Salve counter propagating beam with a beam dump between the two SL injection mirrors (LB_M1 and LB_M2). The ML beam power on the Laser bench was distributed as follows:

  • Power ML fiber output on LB = 263 mW
  • Power at Fiber AMP pick-off = 81 mW
  • Power injected into SL = 182 mW

We switched-off the ML, and then moved it and its driver in the atrium.  We put the same settings: ML Current = 2.000 A, ML Temperature = 22.79°. We optimized the fiber alignment and we could increase the coupling up to 60%. To calculate the coupling we measured the output power at the end of a 2m-fiber connected to the ML box and the input power before the diaphragm inside the box.

The long fiber with the thin metallic jacket that was used to bring the ML beam from the atrium into the laser lab had a damaged end (the one in the laser lab), on saturday the fiber was polished to remove the burned part (55150). Yesterday we cheked that there aren't any important losses by replacing the 2m-fiber used for the ML fiber coupling measurement by this long fiber. The output power was the same as the one measured with the 2m-fiber within 1%, so the fiber is validated.

5 BNC cables were put back to the atrium through the cable entry: EOM_CORR_IN, EOM_CORR_MONIT, ML_TH_CORR, ML_PZT_CORR, ML_PD. The snake with the lemo cables for the monitoring signals of the ML pump diodes is still in the laser lab.

Since the pig-tailed EOM can accept only up to 150 mW of input power, most of the ML power must be dumped, for that an absorbing blade provided by the EGO optics group was installed on the reflection of the OI's input polarizer (see second photo). By turning the half-wave plate after the ML head, the power measured at the end of the 2m-fiber is: 80 mW at 265°, 100 mW at 263°, 120 mW at 261°.

We plugged the 2m ML fiber with 120 mW at the EOM input of the EOM box, at the output we measured 45 mW (using a 1m fiber). We plugged the 300mW fiber AMP input fiber at the EOM output box. As the seeded AMP was off the power measured at the its fiber output was 25 mW. Then we switched-on the fiber AMP, the output power was 314 mW. We connected the long fiber with the metallic jacket. Each time we connected a fiber we checked it with the fiberscope and we often had to clean the fiber end. The bleu fiber connected on the laser bench was damaged and we measured 12% of power losses so we decided to plug directely the long fiber on the laser bench. In order to be sure that we do not misalign the SL input beam by unplugging and replugging the fiber on the collimator, we monitored the output power at the pick-off for the Fiber AMP. After the ML fiber replacement, the pick-off fiber was still well aligned, we optimized by ajusting the mirror LB_M0B. We turned the half-waved plate before the pick-off to send more power to the SL. The power distribution is now:

  • Power ML fiber output on LB = 265 mW
  • Power at Fiber AMP pick-off = 5 mW
  • Power injected into SL = 231 mW

We removed the beam dump from the slave input, it locked without any problems.

 

Images attached to this report
Comments to this report:
Tringali, Derossi, Chiummo - 17:47 Thursday 17 March 2022 (55206) Print this report

After the intervention of the PSL team, the increase of the acoustic (*EIB_MIC, *LB_MIC) and seismic (*EIB_ACC_Y, *LB_ACC_Z, *LB_SEIS*) noise in the Laser Lab observed since Febr 14 in coincidence of the installation of 2 W amplifier (#54833) (Fig.1) and for frequency above 100 Hz, seems to be back to the nominal condition (purple curves), Fig 2.

Images attached to this comment
AdV-INJ (Input Mode Cleaner cavity)
chaibi, chiummo, soulard, turconi - 1:10 Monday 14 March 2022 (55155) Print this report
INJ recovery

After the work on PSL to move the master laser back in the atrium and to restart the Neovan, we recovered the INJ subsystem up to FMODER_tuned state. The arms look rather misaligned, I tried a little but did not manage to find an alignment good enough to engage the lock.

We leave the ITF down, INJ in IMC_RESTORED.

 

AdV-PSL (Laser Bench setup)
chaibi, chiummo, soulard, turconi - 20:16 Saturday 12 March 2022 (55150) Print this report
Polishing of the atrium to the LB fiber
The long fiber, with metallic jacket, which was burned(#51856) was first polished by a 1µm than 0.02 µm texture paper. The fiber seems to be OK but this needs to be confirmed by the transmission measurement.
AdV-PSL (Laser Bench setup)
chaibi, chiummo, soulard, turconi - 19:56 Saturday 12 March 2022 (55149) Print this report
Comment to Swap of the Neovan laser diode driver (55148)
Around 6pm UTC, diodes 1&2 shut down at the same time. After switching OFF then ON the Neovan controller, all diodes were OK. For the sake of security, we choose to switch OFF the Neovan for tonight and swap the picolas diode driver by a new one tomorrow. To be continued...
AdV-PSL (Laser Bench setup)
chaibi, chiummo, soulard, turconi - 19:52 Saturday 12 March 2022 (55148) Print this report
Swap of the Neovan laser diode driver
The old Neovan laser diode driver was responsible of a peak at 1151 Hz measured on the intensity noise of the Neovan amplifier beam (see figure1). The following procedure was followed to swap this diode driver (reference will be given shortly):
- The flip mirror on the PMC transmission was set in the up position and the PMC rampauto lock trigger was set on the PMC_TRA mode.
- The Neovan diodes current was progressively reduced down to zero
- The SL beam was dumped.
- The Neovan controller was switched off
- The new driver was installed and connected.
- The SL was reinjected and the Neovan diodes current was increased to the usual values (4.7A for D1&2 ; 5.3A for D3&4).
The PMC re-locked directly.

We also tried to purge the water circuit of the Neovan diodes :
- The direction of the water flow was reversed for few minutes.
- The derivation to the SL diodes was closed for few seconds (while controlling the T° of the diodes and the Peltier current) in order to increase the flow to the Neovan diodes.
As seen on figure 3&4, it seems that the T° of the diodes in normal operation are 2°C lower.

Images attached to this report
Comments to this report:
chaibi, chiummo, soulard, turconi - 19:56 Saturday 12 March 2022 (55149) Print this report
Around 6pm UTC, diodes 1&2 shut down at the same time. After switching OFF then ON the Neovan controller, all diodes were OK. For the sake of security, we choose to switch OFF the Neovan for tonight and swap the picolas diode driver by a new one tomorrow. To be continued...
derossi, chiummo - 18:19 Sunday 13 March 2022 (55153) Print this report

Due to the failure of the picolas that was installed yesterday (plot 1), this afternoon we replaced the NEOVAN driver of the diodes 1 & 2 (picture 2) with a new one (picture 3).

We took advantage to try to clean the water cooling circuit of the neovan controller. We got rid of some small metallic debris inside the rapid plugs of the water pipes.

We will wait the availability of the beam to restart it.

Images attached to this comment
AdV-PSL (High Power fiber amplifier)
turconi, chaibi, soulard - 19:09 Saturday 12 March 2022 (55147) Print this report
RIN measurements of ALS 300mW fiber amplifiers

Since the restart of PSL/INJ with the neovan amplifier (54833), the master laser is amplified by an additional fiber amplifier after the pig-tailed EOM because the latter cannot stand more than 100 mW. We recall that we want to keep this actuator for the IMC frequency loop in order to prevent fast unlocks.

Two 300mW fibered amplifiers were purchased to Azurlight Systems for the ML amplification: S/N 609 and S/N 610. In the attached figure we show the relative power noise measurements of the two amplifiers compared with the seeder ( NPRO ). The amplifier noise is slightly higher than the seeder noise below 10 kHz. This behaviour is expected and was already observed on the high power amplifier (see VIR-0634A-20 p.5).This additional noise is caused by the pump diodes's power supply. No noise is added at higher frequencies. (NB. Data is not corrected by the photodiode transfert function that is responsible for the noise decrease and the abrut increase visible after 20 MHz).

Tomorrow we will replace the installed keopsys amplifier by one of these amplifiers.

 

 

Images attached to this report
AdV-PSL (Laser Bench setup)
chaibi, gosselin, turconi - 19:19 Wednesday 17 November 2021 (53892) Print this report
ML power

The ML power signal measured on the laser bench (PSL_ML_DC) has decreased since months.
Yesterday we measured the ML fiber output on the laser bench (LB), it was 315 mW while the last reported mesurement was 470 mW (51856). We improve the fiber alignmment and on the ML breadboard and managed to obtain 360 mW  on the LB. 114mW were measured on the pick-off beam for the fiber amplifier seed.

This afternoon we connected back the signal of the photodiode that monitors the ML power directly on its breadboard (PSL_ML_DC_MONIT). This will help in understanding the power drop and decide the next actions.

AdV-PSL (Laser Bench setup)
chaibi, gosselin, turconi - 20:35 Tuesday 16 November 2021 (53881) Print this report
LB Power budget + LB_WP6/7 replacement
This morning we performed some power measurements on the laser bench.
We first switched-off the fiber amplifer and let only the preamp ON(initial current: 7.28 A, power displayed: 98.7W). We also unplugged the PSTAB correction sent to the AOM (NO 1st order diffraction after the AOM).

The average power measured at the laser head output is 1.2 W.
Compared to the power at the laser head level, the losses at different points on the LB are:
- after LB_L7 (before power control at PMC input) : 3%
- after LB_M11 (before PMC input): 7 %
The 3% losses are compatible with the fact that the beam passes through 6 lenses.

4% of the power is lost through 2 half-wave plates, a brewster polarizer and the AOM (from element LB_WP6). We removed then the waveplate before the AOM (LB_WP7). We noticed that it had a foggy surface, with a IR viewer we could observe that the diffused light varies in intensity when changing the position of the beam on the plate suface. Without that waveplate, the measured losses dropped to 2%.

We inspected the other waveplate (LB_WP6) as well. It had a visible spot on its surface. We decided to replace both waveplates. The new ones are from solid photonics (ref: WPO-L/2-1.0-1064, PO: 285/17). See attached picture to locate the waveplate position on the LB.

After the waveplates replacement, the measured losses (at low power) from LB_WP6 to the AOM were still about 4%. Further tests and measurements are required to conclude.

The fiber laser was switched-on again (current: 7.28 A, power displayed: 98.7W), the PMC locked, its alignment optimized. The PSTAB correction was plugged to the AOM. The LB_WP7 that turns the polarization from horizontal to vertical before the AOM was adjusted by optimizing the power at the output of the EIB bench: INJ_EIB_POUT_PD (see attached image 2).

The calibration of the PMC transmission and reflection has changed now because of the LB_WP7 replacement and tuning.




Images attached to this report
AdV-PSL (High Power fiber amplifier)
spinicelli, soulard, turconi - 16:00 Tuesday 31 August 2021 (53018) Print this report
Comment to PSL activities (52967)
On the 25/08/2021 at about 18:00 UTC we switched off the fiber amplifier and switched on the Neovan following this procedure:
- reduce the fiber amp current
- switch-off the fiber preamp
- unscrew and push back the optical rail carrier that carries the fiber amplifier LB input mirror (after the neovan output)
- flip the mirror at the neovan output
- switch on the neovan electronics in the EE room
- increase slowly the pump diodes on the software interface (target: 4.7 A D1/D2 and 5.3 D3/D4)
- release the slave laser beam with the flipping the mirror
- turn LB_WP4 to reach maximum input power in the neovan
- align the PMC with PMC input mirrors (LB_M11, LB_M12)
- remove 9 dBs from the PMC rampeauto AC_IN (PMC error signal for a good lock was about 400 mV pk-pk)

The TEM 00 coupling just after alignment optimization was 82%, the PMC output power was estimated at 61.4W with the PSL_PMC_TRA_DC calibration. At the EIB output, the power was 55.3W (INJ_EIB_POUT_PD). Thus the PMC ouput and the EIB output are very similar (about 1% higher) to the values obtained with the fiber amplifier before the swap.

Then the neovan was switched off and the fiber amplifier was switched on again with 2° colder water in the chillers (see log#52982).

In the figure one can see the AMP diagbox signals together with the PMC reflection, transmission and the EIB output signals around the Neovan-fiber amplifier swap.
Images attached to this comment
AdV-PSL (High Power fiber amplifier)
soulard, turconi - 15:58 Tuesday 31 August 2021 (52988) Print this report
Comment to PSL activities (52967)

Power measurements and DD signals calibration
On the 24/08/2021 we first realigned the beam in the PMC, we could improve the TEM00 coupling from 86% to 87% (measured on PSL_PMC_REFL_DC). This value is the expected one after alignment optimization.
Then we measured the power at the PMC input (PMC IN in the table below) after LB_M12 (last mirror before PMC) for different values of the pump current of the fiber amplifier (I amp). For the same current values we also measured the reflected (PMC_REFL) and transmitted power (PMC_TRA) when the PMC was locked using the PSL_PMC_TRA_DC as threshold. The aim is to calibrate the photodiodes' signals acquired on DataDisplay.


Remarks: For the PMC input and reflected power we blocked the 1st order of the AOM diffraction used for PSTAB (5% of the AOM input power). The transmitted power measurements were done on the 25/08/2021 afternoon. There is a 3% loss between the PMC input power and the power reflected when the PMC is unlocked (see https://logbook.virgo-gw.eu/virgo/?r=41917 ). In the following tables we also show the output power value that one can read on the fiber amplifier display (see log entry https://logbook.virgo-gw.eu/virgo/?r=52982 for power measurements we did at the direct output of the fiber amplifier).
         PSL_AMP_DC_mean Calibration
P displayed (W)   I amp (A)     PMC IN (W)     PSL_AMP_DC (V)  
89                          6.61               75                       0.392    
85                          6.34               70.9                    0.368       
80                          5.99               67.1                    0.349
75                          5.64               63.8                    0.331
70                          5.3                 58.4                    0.304

The calibration is: PMC_IN(W) = 188.96 * PSL_AMP_DC(V) + 1.1696. See data and fit in plot 1


        PSL_PMC_REFL_DC_mean Calibration

P displayed (W)   I amp (A)      PMC REFL (W)      PSL_PMC_REFL_DC (V)   TEM 00 Coupling
89                          6.61                      8.2                         0.09                               87.3%
80                          6.34                      7.7                         0.0839                           87.6%                 
85                          5.99                      7.8                         0.085                             86.7%
80                          5.64                       8                           0.0876                           85%
The TEM00 coupling measurements were done after the power measurements on the PSL_PMC_REFL_DC with the PMC piezo scanning.

The calibration is PMC_REFL(W) = 86.10* PSL_PMC_REFL_DC(V) + 0.49, see plot 2 .


      PSL_PMC_TRA_DC_mean Calibration
 P displayed (W)   I amp (A)      PMC TRA (W)      PSL_PMC_TRA_DC (V)   TEM 00 Coupling
89                          6.61                      61.1                        1.005                               87.4%
80                          6.34                      58.8                        0.965                               86.9%                 
85                          5.99                      55.7                        0.914                               86.9%
80                          5.65                      51.4                        0.843                               86%

The calibration is PMC_TRA(W) = 60.09* PSL_PMC_TRA_DC(V) + 0.75, see plot 3 .

 

Images attached to this comment
AdV-PSL (Characterization of Fiber amplifiers in the lab)
turconi, soulard - 15:05 Friday 27 August 2021 (52982) Print this report
Power measurement at the output of the fiber amplifier and decrease of the chiller temperature to 18°C
We measured the power delivered by the fiber laser on its table (after the telescope) on 25th evening of August
The power delivered by the preamps (I=0A) was 1.5W. It was about 2W (51116) in march 2021.
So, probably the drop of the power at around 80W (I=6.61A) comes from the preamps behaviors.
Then we measured the power of the fiber laser itself leading to Pmeas=14.406*Idisp(A)-12.9 W. for I>1.2A.
We correlated Pmeas with Pmon leading to Pmeas=14,57*Pmon(V) + 11,331 for Pmeas>70W
And Pmeas=-0,4711*Pmon (V)^2+19,347*Pmon(V)-0,5651 W for Pmeas>15W

Then we changed the temperature of chiller 1 (down, Water flow=3.7L/min) and chiller 2 (up, Water flow=4.9L/min) to 18°C.
Before the operation, for I=6.61A, Pmeas was 82.9W (temperature of plateau = 7.13V), PMC_TRA=0.995V, EIB_Pout=54.4W and IMC_TRA=40.7W.
30 mn after the temperature change, it was Pmeas = 80.5W (temperature of plateau = 7.83V), PMC_TRA=0.973V.
On the 2021/08/27, 12h, when temperature is stabilized, we have Pestimated=81.8W, PMC_TRA=1.007V, EIB_Pout=55.3W and IMC_TRA=41W.
AdV-PSL (High Power fiber amplifier)
soulard, turconi - 1:10 Thursday 26 August 2021 (52967) Print this report
PSL activities

During the last two afternoon shifts we did some power measurements on the laser bench. The details of these activities will be given later.

Tonight we let the fiber amplifier ON, the beam is blocked on the laser bench at the PMC output, the PMC is scanning.

Comments to this report:
spinicelli - 10:58 Thursday 26 August 2021 (52971) Print this report

After PSL shift last night, this morning I recover INJ system without any problem. 

Moreover, we switched off both diodes and electronics of NEOVAN amplifier after the tests of last days. 

soulard, turconi - 15:58 Tuesday 31 August 2021 (52988) Print this report

Power measurements and DD signals calibration
On the 24/08/2021 we first realigned the beam in the PMC, we could improve the TEM00 coupling from 86% to 87% (measured on PSL_PMC_REFL_DC). This value is the expected one after alignment optimization.
Then we measured the power at the PMC input (PMC IN in the table below) after LB_M12 (last mirror before PMC) for different values of the pump current of the fiber amplifier (I amp). For the same current values we also measured the reflected (PMC_REFL) and transmitted power (PMC_TRA) when the PMC was locked using the PSL_PMC_TRA_DC as threshold. The aim is to calibrate the photodiodes' signals acquired on DataDisplay.


Remarks: For the PMC input and reflected power we blocked the 1st order of the AOM diffraction used for PSTAB (5% of the AOM input power). The transmitted power measurements were done on the 25/08/2021 afternoon. There is a 3% loss between the PMC input power and the power reflected when the PMC is unlocked (see https://logbook.virgo-gw.eu/virgo/?r=41917 ). In the following tables we also show the output power value that one can read on the fiber amplifier display (see log entry https://logbook.virgo-gw.eu/virgo/?r=52982 for power measurements we did at the direct output of the fiber amplifier).
         PSL_AMP_DC_mean Calibration
P displayed (W)   I amp (A)     PMC IN (W)     PSL_AMP_DC (V)  
89                          6.61               75                       0.392    
85                          6.34               70.9                    0.368       
80                          5.99               67.1                    0.349
75                          5.64               63.8                    0.331
70                          5.3                 58.4                    0.304

The calibration is: PMC_IN(W) = 188.96 * PSL_AMP_DC(V) + 1.1696. See data and fit in plot 1


        PSL_PMC_REFL_DC_mean Calibration

P displayed (W)   I amp (A)      PMC REFL (W)      PSL_PMC_REFL_DC (V)   TEM 00 Coupling
89                          6.61                      8.2                         0.09                               87.3%
80                          6.34                      7.7                         0.0839                           87.6%                 
85                          5.99                      7.8                         0.085                             86.7%
80                          5.64                       8                           0.0876                           85%
The TEM00 coupling measurements were done after the power measurements on the PSL_PMC_REFL_DC with the PMC piezo scanning.

The calibration is PMC_REFL(W) = 86.10* PSL_PMC_REFL_DC(V) + 0.49, see plot 2 .


      PSL_PMC_TRA_DC_mean Calibration
 P displayed (W)   I amp (A)      PMC TRA (W)      PSL_PMC_TRA_DC (V)   TEM 00 Coupling
89                          6.61                      61.1                        1.005                               87.4%
80                          6.34                      58.8                        0.965                               86.9%                 
85                          5.99                      55.7                        0.914                               86.9%
80                          5.65                      51.4                        0.843                               86%

The calibration is PMC_TRA(W) = 60.09* PSL_PMC_TRA_DC(V) + 0.75, see plot 3 .

 

Images attached to this comment
spinicelli, soulard, turconi - 16:00 Tuesday 31 August 2021 (53018) Print this report
On the 25/08/2021 at about 18:00 UTC we switched off the fiber amplifier and switched on the Neovan following this procedure:
- reduce the fiber amp current
- switch-off the fiber preamp
- unscrew and push back the optical rail carrier that carries the fiber amplifier LB input mirror (after the neovan output)
- flip the mirror at the neovan output
- switch on the neovan electronics in the EE room
- increase slowly the pump diodes on the software interface (target: 4.7 A D1/D2 and 5.3 D3/D4)
- release the slave laser beam with the flipping the mirror
- turn LB_WP4 to reach maximum input power in the neovan
- align the PMC with PMC input mirrors (LB_M11, LB_M12)
- remove 9 dBs from the PMC rampeauto AC_IN (PMC error signal for a good lock was about 400 mV pk-pk)

The TEM 00 coupling just after alignment optimization was 82%, the PMC output power was estimated at 61.4W with the PSL_PMC_TRA_DC calibration. At the EIB output, the power was 55.3W (INJ_EIB_POUT_PD). Thus the PMC ouput and the EIB output are very similar (about 1% higher) to the values obtained with the fiber amplifier before the swap.

Then the neovan was switched off and the fiber amplifier was switched on again with 2° colder water in the chillers (see log#52982).

In the figure one can see the AMP diagbox signals together with the PMC reflection, transmission and the EIB output signals around the Neovan-fiber amplifier swap.
Images attached to this comment
AdV-PSL (Power stabilization of HP amplifier)
spinicelli, gosselin, turconi - 9:11 Tuesday 08 June 2021 (52065) Print this report
Comment to Rampeauto PSTAB modification (51953)

As a result of the modification in PSTAB rampauto, we observe the disappearance of the low gain peak @16kHz while a new oscillation @47kHz did raise up after few hours (see picture 1). The origin of this new oscillation is still unclear, as it shouldn't be related to a too low/high gain on the PSTAB loop. Moreover, this peak appears on other non related signals (picture 2) and only when the PSTAB loop is engaged (blue line PSTAB off). Further investigations around the whole power stabilization loop will be done.

To be noted: on Saturday 5th, the oscillation frequency moved towards 44kHz without (apparently) any external intervention (picture 3-5).  

 

Images attached to this comment
AdV-PSL (Power stabilization of HP amplifier)
gosselin, spinicelli, montanari, coulon, turconi - 12:30 Friday 28 May 2021 (51953) Print this report
Rampeauto PSTAB modification

After investigation it appeared that the problem of the low frequency oscillation of the PSTAB already appeared in 2019 (https://logbook.virgo-gw.eu/virgo/?r=47534), the modification mentionned in this log entry had been implemented on the spare rampeauto but we realized it was not on the rampeauto currently working (#1712)
This morning we went in EER to do the modification. It consits in adding two diodes on the M9 component (see picture and scheme attached). In order to be able to come back to the current situation in case of unsuspected problem, we did not added the diodes but substitute the current operational amplifier (M9) with another one having the modification.
We closed the PSTAB and for now it seems to work fine (see figure).
We will monitor its behavior in the coming days.

Images attached to this report
Non-image files attached to this report
Comments to this report:
spinicelli, gosselin, turconi - 9:11 Tuesday 08 June 2021 (52065) Print this report

As a result of the modification in PSTAB rampauto, we observe the disappearance of the low gain peak @16kHz while a new oscillation @47kHz did raise up after few hours (see picture 1). The origin of this new oscillation is still unclear, as it shouldn't be related to a too low/high gain on the PSTAB loop. Moreover, this peak appears on other non related signals (picture 2) and only when the PSTAB loop is engaged (blue line PSTAB off). Further investigations around the whole power stabilization loop will be done.

To be noted: on Saturday 5th, the oscillation frequency moved towards 44kHz without (apparently) any external intervention (picture 3-5).  

 

Images attached to this comment
gosselin, spinicelli - 14:49 Tuesday 08 June 2021 (52069) Print this report

This morning we went to invetigate the problem with the PSTAB.
First we measured the TF, the phase was particularly strange (see picture 1).
We tried to change the gain to see if there could be any effect but we could not come back to a normal behavior. 
We tried to unplugged and reppluged the power supply of the rampeauto and it came back to a relatively normal TF.
We tuned the gain and got a UGF of 64 kHz and a phase margin of 30 degrees (see picture 2). For the same UGF we have less phase margin than what we measured two weeks ago ( https://logbook.virgo-gw.eu/virgo/?r=51912

For now it behaves normally but it seems particularly unstable and will recquire a deeper investigation of the problem. 

We also took advantage of being there to plugg back the out of loop photodiode of the PSTAB to the the ADC and a fast DAC on the perturb input of the rampeauto to inject a line to be able to monitor the gain of the loop.
The hardware is ready for those two canals, we will configure them later with Alain.

Images attached to this comment
coulon - 14:54 Tuesday 08 June 2021 (52070) Print this report
Please measure the +- 15 V at either opamp when there is the trouble.
AdV-PSL (Installation on site)
gosselin, turconi - 18:44 Thursday 29 April 2021 (51579) Print this report
Master laser spare moved to the atrium

Yesterday we installed the ML spare on a breadboard and we coupled the beam into a fiber (picture 1). Coupling is 83% (measurements realized with input power from 200mW up to 500 mW), with a pump current of 2.06 A.

Today we moved the breadboard in the atrium (picture 2).

Images attached to this report
AdV-PSL (Laser Bench setup)
gosselin, derossi, turconi - 22:54 Tuesday 27 April 2021 (51550) Print this report
summary of today PSL activities

The neovan was restarted, its beam is directed on a beam dump, we replaced the beam dump by the powermeter head, the measured output power was 90 W with SL at 17.8W and neovan pump diodes at nominal current. A photodiode has been put temporarly on the leak of the mirror just before the power meter head in order to monitor the neovan power spectrum. This will be used to chaeck that the 1.1kHz line will desappear after the picolas driver replacement.

The neovan was switched-off, the beam dump was put back in place and the fiber amplifier was switched on again.

AdV-PSL (Installation on site)
derossi, gosselin, turconi - 18:25 Tuesday 27 April 2021 (51549) Print this report
Comment to Characterization of the SL pumping diodes (51536)

Today we measured the slave output power after LB_WP4: 17.8 W

The PDH error signal is 3 V pk-pk

AdV-PSL (Laser Bench setup)
gosselin, derossi, turconi - 18:11 Tuesday 27 April 2021 (51548) Print this report
Comment to Noise eater and fast unlocks investigation (51168)

Today while looking at the power spectrum at the output of the neovan we realized that the oscillation at 680 kHz was present (see attached picture). The second picture is the oscillation at 620 kHz when the noise eater is off. 

Doing off/on on the noise eater brought things back to normal.

To be noticed that when switching the INJ off on the 24/04 8:20 UTC the line was not visible on SIB2_RFC_PD1_sample signal. After that we do not have any monitor anymore. There were some works in the laser lab yesterday and today there were some works on the ground distribution. 

Images attached to this comment
AdV-PSL (Installation on site)
gosselin, turconi - 23:03 Monday 26 April 2021 (51536) Print this report
Characterization of the SL pumping diodes

The slave laser pump diodes were replaced on the 28th September 2020  #49528 and on the 31st Decembre the ENEL black out caused a drop in the current #50347.

This afternoon we characterized the diodes' output power as function of the current (values are read on the display of the power supply).The power was measured at the output of a spare multimode fiber. The mesurement is shown in figure 1. The diodes are working fine and the slope is 1.12 W/A for diode 1 and 1.06 W/A for diode 2. Those values are compliant with the measurement done on new diodes in #34756.

We inspected the fibers, the fiber of diode 1 is damaged (see figure 2) while the fiber of diode 2 is clean (see figure 3).

We set the pump current so that the output power is 22W which is the nominal value given in the user manual. Than we adjusted the diode temperature otpimizing the slave laser output power on SL_REFL_PD1_DC. Figure 4 shows the signals on DD, SL_REFL_PD1_DC increased from 0.31V (old current values) to 0.48V (new current values).

old --> new settings are:

Diode 1 Curr: 24.8 A --> 26.6 A

Diode 1 Temp: 26.5 °--> 27°

Diode 2 Curr: 25.8 A --> 28.1 A

Diode 2 Temp: 26.1° --> 27.1°

 

Images attached to this report
Comments to this report:
derossi, gosselin, turconi - 18:25 Tuesday 27 April 2021 (51549) Print this report

Today we measured the slave output power after LB_WP4: 17.8 W

The PDH error signal is 3 V pk-pk

AdV-INJ (Input Mode Cleaner cavity)
chaibi, melo, turconi, merzougui, coulon - 16:43 Tuesday 22 December 2020 (50308) Print this report
IMC at 50W

After the works on PSL, the PMC was locked on the fibered laser.

In order to see the working behaviour of the loops with the new power, we locked the PMC with 1.16 V (pumps 7A, fiber amplifier output power 95W, coupling into the PMC 85% since it was opti;ized for a lower power), giving at the output of IMC_TRA = 50W. We closed all loops (BPC, IMC AA in drift control and RFC) for some minutes in order to see if everything was working properly (see attached figures). All loops closed without any problems. The RFC locked in a wrong mode, but since the intention was just to see if it could lock, we didn't spend too much time on alignment (to leave in this condition we should also adjust the power going to the different photodiodes, as PSTAB, B2, ect.)

After the "proof of concept" we restored the previous power on PMC_TRA (0.967V) and ~42W on IMC_TRA which corresponds for the fiber amplifier to 6A, coupling efficiency on the PMC 88%.

We locked again all the loops (BPC, IMC in full bandwidth, PSTAB and RFC), and left the system in this configuration.

 

Images attached to this report
AdV-PSL (High Power fiber amplifier)
chaibi, turconi - 23:08 Monday 21 December 2020 (50301) Print this report
High power fiber amplifier diagnosis
Today, the fiber amplifier beam was injected into the new PMC. As before, the coupling efficiency didn't exceed 70% and was very dependent of the beam power (40% at high power). Successively, the two polarizers (FA-P1 and FA-P2) were removed and the coupling efficiency reached 80% and was slightly dependent on the amplifier output power as expected. The matching telescope needs to be readjusted to increase the coupling efficiency. The polarizer FA-P1 shall be diagnosed in order to understand its behavior change.
AdV-PSL (Laser Bench setup)
chaibi, merzougui, turconi - 22:57 Monday 21 December 2020 (50299) Print this report
Dismount of the oldPMC mirrors
Today, the old PMC chamber was opened in the class1 clean room. All 4 mirrors were dismounted and packaged to be sent to LMA for diagnosis. No apparent damage were observed on the coatings except that the 1/2'' mirror supported by the PZT seems to be different from the other other 1" mirror. To be investigated. Photos will follow.
AdV-PSL (Assembly Structure Transport Mechanics)
coulon, chabi, turconi - 20:46 Monday 21 December 2020 (50297) Print this report
PMC lock with rampeauto
Today we fixed 2 rampeautos. We obtain an automatic lock in 1/f^4. We adjusted the alignment of the master laser into the EOM.
Now the neovan is injected and the PMC is locked.
AdV-PSL (High Power fiber amplifier)
chaibi, turconi - 12:38 Monday 21 December 2020 (50290) Print this report
Comment to Fiber amplifier diagnosis (50285)

In the picture you can see the optical setup at the output of the fiber amplifier (FA) head on the laser bench.  In order there are:

FA_WP1: quarter waveplate (l/4)

FA_WP2: half waveplate(l/2)

FA_P1: polarizer

FA_WP3+FA_P2: power control half waveplate and polarizer

FA_WP4: half waveplate to turn polarization to s (mirrors' HR coating is for s-pol )

FA_M1 and FA_M2: alignment mirrors for the FA PMC-mode-matching telescope.

 

 

Images attached to this comment
AdV-PSL (Laser Bench setup)
coulon, chaibi, turconi - 20:44 Sunday 20 December 2020 (50286) Print this report
Attempts to lock the new PMC over 1/f4
The rampauto PSL_ELEC_1909_04 (which works under a power supply of +/_ 200V, PA85 high voltage amplifier) locks properly under 1/f (as before) and manually under 1/f4 thanks to the gain increase (bandwidth up to 3.5kHz instead of 1.5kHz) but not automatically. Unfortunately, the PA85 had a failure and must replaced at Nice. Meanwhile, we increased one of the power supplies from 120V to 150V in order to use old rampautos (PA83 limited 150V). We are proceeding with the adaptation of these rampautos so they can be used on the new PMC.
Search Help
×

Warning

×