Environmental Monitoring›Environmental Monitoring #68692

sensors at 1500W

We installed one accelerometer and one microphone inside the 1500W building. 

One meggit accelerometer is placed on the concrete floor, veritcally oriented, close to the INGV gravimeter inside the Fiber Lab: picture 1.

One GRASS AZ46 microphone  is placed in the Bd. hall: picture 2.

3 comments

fiori · 27 May 2026 · 11:55 #69135

Graphs show data of the accelerometer installed (vertical) on the floor next to the Gravimeter. Fig.1 shows that the noise conditions is more or less stable over 24 hours. Figure 2 compares the spectral noise with one accelerometer sensor (vertical) on the CEB floor (near Guralp). The Fiber lab floor is noisier above 10Hz and below 1Hz.  

fiori, kraja · 27 May 2026 · 15:27 #69137

Since May 17 the gravimeter was not reachable from remote. This has been fixed today. Attached are two pictures of the gravimeter (Fig.1) and the laser box (Fig.2).

The issue was with the USB-to-RJ45 adapter, which for some reason was not working properly (the network communication LEDs were off, and the network interface was also disconnected from the console). After unplugging and reconnecting it, the connection was restored, but it went down again a few minutes later. The procedure was repeated, and the connection was re-established. Now, after returning, the machine appears to still be up and running.

fiori · 1 Jun 2026 · 08:28 #69147

Data sheets of the Quantum gravimeters are now available in the Environmental Monitoring Sys > Sensors and Tools > Documentation: https://scientists.virgo-gw.eu/EnvMon/List/Gravimeter/