AdV-TCS (CO2 laser projector)
berni, carbognani - 18:11 Tuesday 13 November 2018 (43563) Print this report
Comment to WI CO2 laser chiller failure (43560)

To have back the signal in the DAQ (and then in the DMS) we have:

- replaced the application for TCSChillerWI from /virgoApp/PyArtic/v1r1/scripts/PyArtic.py to /virgoDev/Chiller/v3r4/Linux-x86_64-SL6/ChillerApp.exe in the VPM;

- renamed the old configuration file from TCSChillerAux.cfg to TCSChillerWI.cfg;

- renamed the old channels name with the current ones (from TE_Aux_Chiller to TE_WI_Chiller, from TE_Aux_Chiller_Set to TE_Aux_Chiller_Set, from Aux_Chiller_Alarm to Aux_Chiller_Alarm);

- configured the serial bridge with the proper settings: the device has been connected on port 3 of ethadvtcs2 with 9600 Baudrate;

- properly switched off the TCSChillerWI server;

- restart the process from VPM with the new configurations.