The digital noise correction was not propagated to the DSP, the numerical noise is always computed to monitor this issue.
At 017-03-24-04h25m11-UTC I enable the numiracal noise substraction(NNS)
017-03-24-04h25m11-UTC info masserot 'NNS subtraction:Enable NNS [ramp duration=1,weigh of digital noise corr=1]' sent to SSFS_Ctrl
first plot show the digital noise without the NNS,
second plot with the NNS ON where one can see the reduction of the remaining numerical noise
At 2017-03-24-04h44m03-UTC I reset the DSP filter registers
2017-03-24-04h44m03-UTC info masserot 'SSFS filter:B4: reset all DSP filters [filter=2]' sent to SSFS_Ctrl
third plot with NNS ON show the increase of the digital noise correction after 20mn
the fourth plot with NNS ON show the reduction of the digital noise after reseting the DSP filter registers