ITF found in Science mode.
At 7:00 UTC ITF in Maintenance mode, below the list of the activities communicated in control room:
- standard tasks performed by the operator:
- scan and lock of the OMC in single bounce;
- TCS thermo camera images reference;
- check of the TCS powers:
| CH [W] | OUTER RING [W] | INNER RING [W] | |
| WI | 0.264 | 0.295 | 0.032 |
| NI | 0.682 | 0.57 | 0.569 |
- CRAL building digging work;
- from 9:30 UTC to 10:30 UTC PSTAb realignment in the AOM (INJ team);
- SNEB microsprings and DMS warning;
- Tests of ACS INJ; for this test I set the UTA INJ in PORTATA NOMINALE from 10:13 UTC to 10:49 UTC;
All the activities completed at 11:18 UTC; I manually realigned the cavities, then the CITF and then I adjusted MICH offset. Finally the ITF was in Science mode at 12:40 UTC.
SUSP
This morning the susp on-call tried to fix the high voltage for F0 OB COIL H2; unfurtunately he discovered a bug in the software that is preventing to restore the situation. At the moment the SUSP team is trying to fix the bug and then they will be able to restore the corrections.