AdV-TCS (Ring Heater)
rocchi - 19:20 Wednesday 01 April 2020 (48889) Print this report
RHs temperature sensors
While checking the TCS status, RHs in particular, I noticed the strange behavior of the temperatures measured by all the sensors installed on the TM RH shields, as shown in figure. Note that the ITM RHs are both off. It would be interesting to track what piece of electronics (out of the many switched off yesterday) is interfering with the RH termometers.

The two small drops, clearly visible on the NE_RH, are, as usual, due to the unlocks of the ITF, in this case, those happened on March 30th.
Images attached to this report
Comments to this report:
rocchi, pasqualetti - 22:27 Wednesday 01 April 2020 (48891) Print this report
Thanks to a quick e-mail interaction with Antonio, the issue is clear: tube valves closed obscure the cryotraps from the TMs view.