22:00 UTC, ITF found in Commissioning Mode with SNEB LC still opened. The recovery lasted until 23:40 UTC then, with the help of Vardaro and Bersanetti from remote, we started again to relock ITF.
ITF Locked in LN3 SQZ at 00:25 UTC. Science Mode set at 00:38 UTC with BNS Range 49 Mpc.
ITF unlocked (TBC) at 02:11 UTC. Relocked at 02:55 UTC,
GRB
Fermi event created: E361413 at 3:42 UTC. Alert reset at 03:44 UTC.
Guard Tour (UTC)
23:00 - 23:35
01:00 - 01:35
03:05 - 03:40
05:00 - 05:35
ISC
Fine tuning:
- Vardaro tuned the phase of B4 56 Mhz from -1.51 to -1.24. (Adjusting Mode set)
- DIFFpTy tuning from -0.034 to -0.024. After a while, I tuned again to -0.034.
- DIFFpTx tuning from -0.003 to 0.01
SBE
23:00 UTC I went at WE to switch OFF the 4 HEPA filter above SWEB clean area.
The flag SNEB_B7_Cam1_FitPosY became red. To restore the nominal value I manually tuned the value of SNEB_LC_TX_SET. Value changed from -668 to -678
The flag SNEB_B7_Cam1_FitPosX became yellow. To restore the nominal value I manually tuned the value of SNEB_LC_TY_SET. Value changed from -114 to -110.
ISC
(27-01-2020 21:15 - 28-01-2020 01:00) From remote
Status: Ended
Description: Problem with the lock of the North Arm.