AdV-AEI Squeezer Integration (external squeezer table)De Laurentis, Tacca, Sorrentino - 11:12 Tuesday 20 February 2018 (40636)
Print this reportSqueezer Out Beam CharacterizationYesterday, Monday 19th Feb, we characterized again the squeezer out beam with two different beam profilers to be sure to avoid calibration errors.
We used the THORLABS Scanning Slit BP and the OPHIR Spiricon Camera BP.
The obtained results are:
THORLABS with Sensor at 0 ° (fig. 1):
wox = 0.434 +/- .001 mm
woy = 0.391 +/- .004 mm
zox= 1150 +/- 20 mm
zoy= 1064+/- 6 mm
THORLABS with Sensor at 45 ° (fig. 2):
wox = 0.41 +/- .01 mm
woy = 0.42 +/- .01 mm
zox= 1108 +/-9 mm
zoy= 1140+/- 20 mm
THORLABS with Sensor at -45 °(fig. 3):
wox = 0.41 +/- .01 mm
woy = 0.42 +/- .01 mm
zox= 1101 +/-7 mm
zoy= 1130 +/- 30 mm
SPIRICON (fig. 4):
wox= 0.49 +/- 0.02 mm
woy= 0.42 +/- 0.01 mm
zox = 1240 +/- 30 mm
zoy = 1100 +/- 20 mm
The beam seems to be not a perfect Gaussian, as can be observed in pictures of figures 5 and 6 taken with the SPIRICON camera.
Images attached to this report
Comments to this report:
delauren - 18:19 Tuesday 20 February 2018 (40640)
Print this reportATTENTION: the xlabel of the figures 1-4 is wrong. The right one is 'Dstance from the Squeezer BOX out'